D3711
Troubleshooting Scale Systems Worksheet
With Scale indicator “On”, press [On] to go into self-test. Write down the setup and calibration
numbers for reference. Setup ___________. Calibration ___________.
1. Disconnect the Junction box from the scale indicator. Watch what happens to the
weight. The weight should decrease as the junction box is disconnected. You should
see a negative weight on the display. If the weight does not change when the junction
box is disconnected there is something wrong with the scale indicator.
2. Press [Net/gross] and [Zero] to zero balance the scale with the J-box disconnected.
The weight should stay at zero. If weight is unstable, check scale connector for
moisture. This is your zero reference for the rest of the testing. No need to re-zero
between load cell tests.
3. Re-connect the J-box and see what happens. Note the weight on the display
____________. Should see a positive weight It should be the weight of the mixer,
cart, etc plus whatever weight is in it. If scale indicator says +range, and the weight is at
capacity check setup number to make sure gain setting is correct. A +range or -range
could mean a bad load cell.
4. Open up the junction box. Check for any loose wires, or any wires with insulation in the
connector instead of bare wire. To start the load cell troubleshooting, disconnect all
load cells from the junction box. All 5 colored wires of all load cells have to be
disconnected. Leave the cable to the indicator connected.
5. What is the weight on the display? _____________. It should be within a few counts of
the zero we had in step 2. If not there may be a problem with the junction box cable or
junction box board. Disconnect J-box cable from junction box to see if problem is with
cable or junction box. Weight should be stable, if not check for any moisture in the
junction box, or check cable for any cuts or pinched spots.
6. Next check each load cell. Connect back up the 5 colored wires of one load cell.
Weight should be a positive stable weight. When weight is added over the load cell
that is connected it should go up 3 to 4 times the actual weight. Tap on the load cell to
see if weight moves more than a count or two. Disconnect that load cell and test the
next one.
7. Note the weight on the display when each cell is first connected up in J-box.
1.______________ 2.______________ 3._______________ 4.________________
and weight on the display goes to when weight is added over each cell.
1.______________ 2.______________ 3._______________ 4.________________
Each load cell should go up the same amount. If one load cell does not respond to weight
changes it could be bad. +range, -range or unstable weights also mean a bad load cell.